ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.

Reported: June 28, 2017 Initiated: April 18, 2017 #Z-2613-2017 7981 kits units

Abbott Laboratories issued this FDA Devices recall on June 28, 2017. Classified as Moderate severity (Class II). Approximately 7981 kits units are affected. The recall was issued because: Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a …. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2613-2017) was formally reported on June 28, 2017, with the manufacturer initiating the action on April 18, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Abbott Laboratories is listed as the recalling firm, operating out of Abbott Park, IL. Federal records indicate 7981 kits units are affected.

The documented reason for this recall is: Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBI…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 9 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

7981 kits

Related Recalls

6

6 from same agency

Product Description

ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.

Reason for Recall

Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.

Details

Recalling Firm
Abbott Laboratories
Units Affected
7981 kits
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM
Location
Abbott Park, IL

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2613-2017
Date reported June 28, 2017
Date initiated April 18, 2017
Recalling firm Abbott Laboratories
Units affected 7981 kits
Distribution Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROAT…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

7981 kits units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.. Recalled by Abbott Laboratories. Units affected: 7981 kits.
Why was this product recalled?
Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 28, 2017. Severity: Moderate. Recall number: Z-2613-2017.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2613-2017) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).