Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds

Reported: September 27, 2023 Initiated: August 16, 2023 #Z-2614-2023 38,033 devices units

Abbott Medical issued this FDA Devices recall on September 27, 2023. Classified as Moderate severity (Class II). Approximately 38,033 devices units are affected. The recall was issued because: Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization T…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-2614-2023) was formally reported on September 27, 2023, with the manufacturer initiating the action on August 16, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Abbott Medical is listed as the recalling firm, operating out of Sylmar, CA. Federal records indicate 38,033 devices units are affected.

The documented reason for this recall is: Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of cryst… Distribution data in the federal record shows the product reached: U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lith…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 3 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

38,033 devices

Related Recalls

6

6 from same agency

Product Description

Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV. Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.

Reason for Recall

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

Details

Recalling Firm
Abbott Medical
Units Affected
38,033 devices
Distribution
U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom. Latin America - Pending
Location
Sylmar, CA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-2614-2023
Date reported September 27, 2023
Date initiated August 16, 2023
Recalling firm Abbott Medical
Units affected 38,033 devices
Distribution U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, M…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

38,033 devices units affected — limited or regional distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units) ✓ This recall
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV. Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.. Recalled by Abbott Medical. Units affected: 38,033 devices.
Why was this product recalled?
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 27, 2023. Severity: Moderate. Recall number: Z-2614-2023.
Where was the recalled product distributed?
Distribution: U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom. Latin America - Pending.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2614-2023) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).