The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains …
Becton Dickinson & Co. recalled The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic… - a moderate-severity action.
The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic… was recalled by Becton Dickinson & Co. in September 17, 2014. Reason: The performance of the assay no longer conforms to this statement in the Analytical Specificity section of th…. Check the official notice for the remedy. Verify recall #Z-2618-2014 with the FDA Devices before acting.
The recall
Becton Dickinson & Co. issued this moderate-severity FDA Devices recall-The performance of the assay no longer conforms to this statement in the Analytical Specificity section of th….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2618-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2618-2014) was formally reported on September 17, 2014, with the manufacturer initiating the action on July 23, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Becton Dickinson & Co. is listed as the recalling firm, operating out of Sparks, MD. Federal records list the affected scope as 585,048 tests.
The documented reason for this recall is: The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations … Distribution data in the federal record shows the product reached: US Distribution including the states of : CA, FL, HI, IL, IN, MD, MI, MN, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
585,048 tests
Related Recalls
6
3 from same agency
The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.
The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2618-2014 |
| Date reported | September 17, 2014 |
| Date initiated | July 23, 2014 |
| Recalling firm | Becton Dickinson & Co. |
| Firm location | Sparks, MD |
| Affected scope | 585,048 tests |
| Distribution | US Distribution including the states of : CA, FL, HI, IL, IN, MD, MI, MN, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 17, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.