Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Reported: October 1, 2025 Initiated: August 6, 2025 #Z-2636-2025

Product Description

Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Reason for Recall

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

Details

Units Affected
98 units
Distribution
US Nationwide distribution in the states of FL, IL, MD & TX.
Location
Wayne, NJ

Frequently Asked Questions

What product was recalled?
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.. Recalled by Maquet Cardiovascular, LLC. Units affected: 98 units.
Why was this product recalled?
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 1, 2025. Severity: Moderate. Recall number: Z-2636-2025.