Severity
Moderate
FDA Devices recall · Reported October 4, 2023
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Ste…
3M Company - Health Care Business recalled 3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor… - a moderate-severity action.
3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor… was recalled by 3M Company - Health Care Business in October 4, 2023. Reason: Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-P…. Check the official notice for the remedy. Verify recall #Z-2637-2023 with the FDA Devices before acting.
The recall
3M Company - Health Care Business issued this moderate-severity FDA Devices recall-Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-P….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2637-2023 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2637-2023) was formally reported on October 4, 2023, with the manufacturer initiating the action on August 22, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. 3M Company - Health Care Business is listed as the recalling firm, operating out of Saint Paul, MN. Federal records list the affected scope as 79500 devices.
The documented reason for this recall is: Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biologi… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
79500 devices
Related Recalls
6
3 from same agency
3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2637-2023 |
| Date reported | October 4, 2023 |
| Date initiated | August 22, 2023 |
| Recalling firm | 3M Company - Health Care Business |
| Firm location | Saint Paul, MN |
| Affected scope | 79500 devices |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 4, 2023. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.