ZAP-X Radiosurgery System, Model: 300150

Reported: October 8, 2025 Initiated: July 25, 2025 #Z-2639-2025

Product Description

ZAP-X Radiosurgery System, Model: 300150

Reason for Recall

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Details

Recalling Firm
Zap Surgical Systems
Units Affected
17
Distribution
Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
Location
San Carlos, CA

Frequently Asked Questions

What product was recalled?
ZAP-X Radiosurgery System, Model: 300150. Recalled by Zap Surgical Systems. Units affected: 17.
Why was this product recalled?
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.