FDA Devices recall · Reported October 8, 2025

VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Recall #
Z-2671-2025
Affected scope
235269 cards
Initiated
June 26, 2025
Verify with FDA Devices →
View my saved recalls

bioMerieux, Inc. recalled VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VI… - a moderate-severity action.

VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VI… was recalled by bioMerieux, Inc. in October 8, 2025. Reason: Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa a…. Check the official notice for the remedy. Verify recall #Z-2671-2025 with the FDA Devices before acting.

The recall

bioMerieux, Inc. issued this moderate-severity FDA Devices recall-Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa a….

Moderate
severity level
Class II
classification
October 8, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2671-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only

Reason for recall

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-2671-2025
Date reported October 8, 2025
Date initiated June 26, 2025
Recalling firm bioMerieux, Inc.
Firm location Durham, NC
Affected scope 235269 cards
Distribution International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Mace…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2671-2025) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only. Recalled by bioMerieux, Inc.. Units affected: 235269 cards.
Why was this product recalled?
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2025. Severity: Moderate. Recall number: Z-2671-2025.
Where was the recalled product distributed?
Distribution: International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Poland, Portugal, Romania, Russia, Serbia, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkey..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported October 8, 2025. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026