Severity
Moderate
FDA Devices recall · Reported August 28, 2024
1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red bl…
Abbott Laboratories recalled Alinity hq Analyzer, REF: 09P68-01 - a moderate-severity action.
Alinity hq Analyzer, REF: 09P68-01 was recalled by Abbott Laboratories in August 28, 2024. Reason: 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately …. Check the official notice for the remedy. Verify recall #Z-2681-2024 with the FDA Devices before acting.
The recall
Abbott Laboratories issued this moderate-severity FDA Devices recall-1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2681-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2681-2024) was formally reported on August 28, 2024, with the manufacturer initiating the action on June 14, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Abbott Laboratories is listed as the recalling firm, operating out of Santa Clara, CA. Federal records list the affected scope as 816.
The documented reason for this recall is: 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating fal… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CH…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
816
Related Recalls
6
3 from same agency
Alinity hq Analyzer, REF: 09P68-01
1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2681-2024 |
| Date reported | August 28, 2024 |
| Date initiated | June 14, 2024 |
| Recalling firm | Abbott Laboratories |
| Firm location | Santa Clara, CA |
| Affected scope | 816 |
| Distribution | Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROAT… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 28, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.