FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System

Reported: September 4, 2024 Initiated: July 24, 2024 #Z-2692-2024

Product Description

FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System

Reason for Recall

Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.

Details

Units Affected
8,174
Distribution
US distribution to states of: NJ, PA, OH, TX, VA, CA, NY, FL, WA, MS, MN, and WY
Location
Alameda, CA

Frequently Asked Questions

What product was recalled?
FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Recalled by Abbott Diabetes Care, Inc.. Units affected: 8,174.
Why was this product recalled?
Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 4, 2024. Severity: Critical. Recall number: Z-2692-2024.