BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.

Reported: September 7, 2016 Initiated: August 8, 2016 #Z-2710-2016

Product Description

BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.

Reason for Recall

Becton Dickinson is recalling various lot numbers of BD Vacutainer Push Button Blood Collection Set because of reports that specific lots exhibit separation of fron and rear barrels upon activation of the safety feature, which retracts the needle.

Details

Units Affected
1,788,150 units
Distribution
US and International
Location
Franklin Lakes, NJ

Frequently Asked Questions

What product was recalled?
BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.. Recalled by Becton Dickinson & Company. Units affected: 1,788,150 units.
Why was this product recalled?
Becton Dickinson is recalling various lot numbers of BD Vacutainer Push Button Blood Collection Set because of reports that specific lots exhibit separation of fron and rear barrels upon activation of the safety feature, which retracts the needle.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 7, 2016. Severity: Moderate. Recall number: Z-2710-2016.