FDA Devices recall · Reported October 8, 2014

Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.

Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension V…

Recall #
Z-2720-2014
Affected scope
9769 Total (6000 US)
Initiated
August 18, 2014
Verify with FDA Devices →
View my saved recalls

Siemens Healthcare Diagnostics, Inc. recalled Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnos… - a moderate-severity action.

Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnos… was recalled by Siemens Healthcare Diagnostics, Inc. in October 8, 2014. Reason: Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC…. Check the official notice for the remedy. Verify recall #Z-2720-2014 with the FDA Devices before acting.

The recall

Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC….

Moderate
severity level
Class II
classification
October 8, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2720-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.

Reason for recall

Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2720-2014
Date reported October 8, 2014
Date initiated August 18, 2014
Recalling firm Siemens Healthcare Diagnostics, Inc.
Firm location Newark, DE
Affected scope 9769 Total (6000 US)
Distribution Worldwide Distribution - USA (Nationwide) and Internationally to Australia Austria Belgium Canada Czech Republic Finland France Germany Great Britain Italy Japan Netherlands New Zealand Norway Portugal Slovak Republic Slove…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2720-2014) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 9769 Total (6000 US).
Why was this product recalled?
Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2014. Severity: Moderate. Recall number: Z-2720-2014.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA (Nationwide) and Internationally to Australia Austria Belgium Canada Czech Republic Finland France Germany Great Britain Italy Japan Netherlands New Zealand Norway Portugal Slovak Republic Slovenia South Korea Spain Switzerland.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported October 8, 2014. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026