Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.

Reported: October 8, 2014 Initiated: August 18, 2014 #Z-2720-2014

Product Description

Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.

Reason for Recall

Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.

Details

Units Affected
9769 Total (6000 US)
Distribution
Worldwide Distribution - USA (Nationwide) and Internationally to Australia Austria Belgium Canada Czech Republic Finland France Germany Great Britain Italy Japan Netherlands New Zealand Norway Portugal Slovak Republic Slovenia South Korea Spain Switzerland
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista¿ System.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 9769 Total (6000 US).
Why was this product recalled?
Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2014. Severity: Moderate. Recall number: Z-2720-2014.