Severity
Critical
FDA Devices recall · Reported August 12, 2020
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infu…
CareFusion 303, Inc. recalled Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system - a critical-severity action.
Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system was recalled by CareFusion 303, Inc. in August 12, 2020. Reason: Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communica…. Check the official notice for the remedy. Verify recall #Z-2721-2020 with the FDA Devices before acting.
The recall
CareFusion 303, Inc. issued this critical-severity FDA Devices recall-Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communica….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2721-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2721-2020) was formally reported on August 12, 2020, with the manufacturer initiating the action on June 30, 2020. It is classified under Critical severity (Class I), with a current status of Terminated. CareFusion 303, Inc. is listed as the recalling firm, operating out of San Diego, CA. Federal records list the affected scope as 38982.
The documented reason for this recall is: Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNEC… Distribution data in the federal record shows the product reached: Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, D…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
38982
Related Recalls
6
3 from same agency
Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-2721-2020 |
| Date reported | August 12, 2020 |
| Date initiated | June 30, 2020 |
| Recalling firm | CareFusion 303, Inc. |
| Firm location | San Diego, CA |
| Affected scope | 38982 |
| Distribution | Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, a… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 12, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.