Severity
Moderate
FDA Devices recall · Reported September 14, 2016
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caus…
Philips Medical Systems (Cleveland) Inc recalled BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography … - a moderate-severity action.
BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography … was recalled by Philips Medical Systems (Cleveland) Inc in September 14, 2016. Reason: The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, …. Check the official notice for the remedy. Verify recall #Z-2736-2016 with the FDA Devices before acting.
The recall
Philips Medical Systems (Cleveland) Inc issued this moderate-severity FDA Devices recall-The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2736-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2736-2016) was formally reported on September 14, 2016, with the manufacturer initiating the action on May 25, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems (Cleveland) Inc is listed as the recalling firm, operating out of Cleveland, OH. Federal records list the affected scope as 424 units.
The documented reason for this recall is: The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow … Distribution data in the federal record shows the product reached: Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republi…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
424 units
Related Recalls
6
3 from same agency
BrightView XCT Model 882482, gamma camera for Single Photon Emission Computed Tomography (SPECT)
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2736-2016 |
| Date reported | September 14, 2016 |
| Date initiated | May 25, 2016 |
| Recalling firm | Philips Medical Systems (Cleveland) Inc |
| Firm location | Cleveland, OH |
| Affected scope | 424 units |
| Distribution | Worldwide distribution. US nationwide, Algeria, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Finland France, Gabon, Germany, Greece, India, Indonesia, Ireland, Japan, Korea, Republic of Korea, Republic… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 14, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.