Severity
Critical
FDA Devices recall · Reported August 2, 2017
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient…
Maquet Datascope Corp - Cardiac Assist Division recalled CS 300 Intra-Aortic Balloon Pump - a critical-severity action.
CS 300 Intra-Aortic Balloon Pump was recalled by Maquet Datascope Corp - Cardiac Assist Division in August 2, 2017. Reason: The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance c…. Check the official notice for the remedy. Verify recall #Z-2738-2017 with the FDA Devices before acting.
The recall
Maquet Datascope Corp - Cardiac Assist Division issued this critical-severity FDA Devices recall-The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance c….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2738-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2738-2017) was formally reported on August 2, 2017, with the manufacturer initiating the action on June 16, 2017. It is classified under Critical severity (Class I), with a current status of Terminated. Maquet Datascope Corp - Cardiac Assist Division is listed as the recalling firm, operating out of Mahwah, NJ. Federal records list the affected scope as 12,319 units total.
The documented reason for this recall is: The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device … Distribution data in the federal record shows the product reached: Worldwide Distribution - (in over 100 countries) US (Nationwide). Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
12,319 units total
Related Recalls
6
3 from same agency
CS 300 Intra-Aortic Balloon Pump
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-2738-2017 |
| Date reported | August 2, 2017 |
| Date initiated | June 16, 2017 |
| Recalling firm | Maquet Datascope Corp - Cardiac Assist Division |
| Firm location | Mahwah, NJ |
| Affected scope | 12,319 units total |
| Distribution | Worldwide Distribution - (in over 100 countries) US (Nationwide) |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 2, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.