Severity
Moderate
FDA Devices recall · Reported September 14, 2016
Data from internal studies suggests there may be a potential that the secondary package seal integrity does not provide the level of sterility intended for Minimally Invasive Achi…
Integra LifeSciences Corp. recalled Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Su… - a moderate-severity action.
Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Su… was recalled by Integra LifeSciences Corp. in September 14, 2016. Reason: Data from internal studies suggests there may be a potential that the secondary package seal integrity does n…. Check the official notice for the remedy. Verify recall #Z-2743-2016 with the FDA Devices before acting.
The recall
Integra LifeSciences Corp. issued this moderate-severity FDA Devices recall-Data from internal studies suggests there may be a potential that the secondary package seal integrity does n….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2743-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2743-2016) was formally reported on September 14, 2016, with the manufacturer initiating the action on June 17, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Integra LifeSciences Corp. is listed as the recalling firm, operating out of Plainsboro, NJ. Federal records list the affected scope as 6,729 units.
The documented reason for this recall is: Data from internal studies suggests there may be a potential that the secondary package seal integrity does not provide the level of sterility intended for Minimally Invasive Achilles Tendon Suture System Achillon. Distribution data in the federal record shows the product reached: Nationwide distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
6,729 units
Related Recalls
6
3 from same agency
Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment The Achillon Suture System is a single use, sterile device used to treat acute Achilles tendon ruptures. The Achillon Suture System is a mini-invasive procedure indicated to teat acute Achilles tendon ruptures: Acute rupture of the Achill tendon(<10 days); Rupture located between 2 and 8 cm above calcaneum; Open or closed ruptures
Data from internal studies suggests there may be a potential that the secondary package seal integrity does not provide the level of sterility intended for Minimally Invasive Achilles Tendon Suture System Achillon.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2743-2016 |
| Date reported | September 14, 2016 |
| Date initiated | June 17, 2016 |
| Recalling firm | Integra LifeSciences Corp. |
| Firm location | Plainsboro, NJ |
| Affected scope | 6,729 units |
| Distribution | Nationwide distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 14, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.