FDA Devices recall · Reported August 22, 2018
H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automate…
Terumo Cardiovascular Systems Corporation recalled CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The C… - a moderate-severity action.
CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The C… was recalled by Terumo Cardiovascular Systems Corporation in August 22, 2018. Reason: H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an incr…. Check the official notice for the remedy. Verify recall #Z-2750-2018 with the FDA Devices before acting.
The recall
Terumo Cardiovascular Systems Corporation issued this moderate-severity FDA Devices recall-H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an incr….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2750-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2750-2018) was formally reported on August 22, 2018, with the manufacturer initiating the action on June 26, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Terumo Cardiovascular Systems Corporation is listed as the recalling firm, operating out of Ann Arbor, MI. Federal records list the affected scope as 128 in total.
The documented reason for this recall is: H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered o… Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, SC, TN, TX, WA, and WV. Foreign distribution to Middle East, Latin America, Europe, Japan, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
128 in total
Related Recalls
6
3 from same agency
CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.
H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2750-2018 |
| Date reported | August 22, 2018 |
| Date initiated | June 26, 2018 |
| Recalling firm | Terumo Cardiovascular Systems Corporation |
| Firm location | Ann Arbor, MI |
| Affected scope | 128 in total |
| Distribution | Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, SC, TN, TX, WA, and WV. Foreign distribution to Middle East, Latin America, Europe, Japan, Canada, and Mexico. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 22, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.