FDA Devices recall · Reported July 29, 2026
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may …
Ferno-Washington Inc recalled Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intende… - a moderate-severity action.
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intende… was recalled by Ferno-Washington Inc in July 29, 2026. Reason: Product may experience accelerated roller wear and potential deformation due to interference between the roll…. Check the official notice for the remedy. Verify recall #Z-2760-2026 with the FDA Devices before acting.
The recall
Ferno-Washington Inc issued this moderate-severity FDA Devices recall-Product may experience accelerated roller wear and potential deformation due to interference between the roll….
- Moderate
- severity level
- 278 units
- affected scope
- Class II
- classification
- July 29, 2026
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-2760-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This FDA Devices action (record #Z-2760-2026) was formally reported on July 29, 2026, with the manufacturer initiating the action on June 19, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Ferno-Washington Inc is listed as the recalling firm, operating out of Wilmington, OH. Federal records list the affected scope as 278.
The documented reason for this recall is: Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movem… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Net…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Reason for recall
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2760-2026 |
| Date reported | July 29, 2026 |
| Date initiated | June 19, 2026 |
| Recalling firm | Ferno-Washington Inc |
| Firm location | Wilmington, OH |
| Affected scope | 278 |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Ge… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-2760-2026) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026