PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported October 9, 2019

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Recall #
Z-2763-2019
Affected scope
Access Cortisol Reagent Assay
Initiated
September 18, 2018
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Beckman Coulter Inc. recalled The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic part… - a moderate-severity action.

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic part… was recalled by Beckman Coulter Inc. in October 9, 2019. Reason: Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient…. Check the official notice for the remedy. Verify recall #Z-2763-2019 with the FDA Devices before acting.

The recall

Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient….

Moderate
severity level
Class II
classification
October 9, 2019
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2763-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-2763-2019) was formally reported on October 9, 2019, with the manufacturer initiating the action on September 18, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Beckman Coulter Inc. is listed as the recalling firm, operating out of Brea, CA. Federal records list the affected scope as Access Cortisol Reagent Assay.

The documented reason for this recall is: Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%). Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

Access Cortisol Reagent Assay

Related Recalls

6

3 from same agency

Product description

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Reason for recall

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2763-2019
Date reported October 9, 2019
Date initiated September 18, 2018
Recalling firm Beckman Coulter Inc.
Firm location Brea, CA
Affected scope Access Cortisol Reagent Assay
Distribution Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2763-2019) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.. Recalled by Beckman Coulter Inc.. Units affected: Access Cortisol Reagent Assay.
Why was this product recalled?
Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
Which agency issued this recall?
This recall was issued by the FDA Devices on October 9, 2019. Severity: Moderate. Recall number: Z-2763-2019.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-2763-2019) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 9, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.