Severity
Moderate
FDA Devices recall · Reported September 30, 2015
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching l…
Leica Microsystems, Inc. recalled The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica AS… - a moderate-severity action.
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica AS… was recalled by Leica Microsystems, Inc. in September 30, 2015. Reason: The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort …. Check the official notice for the remedy. Verify recall #Z-2804-2015 with the FDA Devices before acting.
The recall
Leica Microsystems, Inc. issued this moderate-severity FDA Devices recall-The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2804-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2804-2015) was formally reported on September 30, 2015, with the manufacturer initiating the action on July 27, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Leica Microsystems, Inc. is listed as the recalling firm, operating out of Buffalo Grove, IL. Federal records list the affected scope as U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems..
The documented reason for this recall is: The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 … Distribution data in the federal record shows the product reached: Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to U…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems.
Related Recalls
6
3 from same agency
The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.
The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2804-2015 |
| Date reported | September 30, 2015 |
| Date initiated | July 27, 2015 |
| Recalling firm | Leica Microsystems, Inc. |
| Firm location | Buffalo Grove, IL |
| Affected scope | U.S. Quantity Distributed: 163 systems; Foreign Quantity Diistributed: 410 systems. |
| Distribution | Worldwide Distribution - US including AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH,PA, SC, TN, TX, VA, WA, WI, WV and District of Columbia. Foreign distribution to United Arab Emirates,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 30, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.