Severity
Low
FDA Devices recall · Reported September 21, 2016
There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.
Ortho-Clinical Diagnostics recalled The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITR… - a low-severity action.
The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITR… was recalled by Ortho-Clinical Diagnostics in September 21, 2016. Reason: There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a succ…. Check the official notice for the remedy. Verify recall #Z-2804-2016 with the FDA Devices before acting.
The recall
Ortho-Clinical Diagnostics issued this low-severity FDA Devices recall-There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a succ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2804-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2804-2016) was formally reported on September 21, 2016, with the manufacturer initiating the action on April 19, 2016. It is classified under Low severity (Class III), with a current status of Terminated. Ortho-Clinical Diagnostics is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as 520 Units.
The documented reason for this recall is: There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay. Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified low severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Low
Affected scope
520 Units
Related Recalls
6
3 from same agency
The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.
There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Low (Class III) |
| Status | Terminated |
| Recall number | Z-2804-2016 |
| Date reported | September 21, 2016 |
| Date initiated | April 19, 2016 |
| Recalling firm | Ortho-Clinical Diagnostics |
| Firm location | Rochester, NY |
| Affected scope | 520 Units |
| Distribution | Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colomb… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 21, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.