LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging

Reported: September 2, 2020 Initiated: January 31, 2020 #Z-2820-2020

Product Description

LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging

Reason for Recall

The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).

Details

Recalling Firm
89 NORTH
Units Affected
188
Distribution
Worldwide
Location
Williston, VT

Frequently Asked Questions

What product was recalled?
LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging. Recalled by 89 NORTH. Units affected: 188.
Why was this product recalled?
The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).
Which agency issued this recall?
This recall was issued by the FDA Devices on September 2, 2020. Severity: Moderate. Recall number: Z-2820-2020.