Severity
Moderate
FDA Devices recall · Reported September 2, 2020
The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode inten…
89 North recalled LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 … - a moderate-severity action.
LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 … was recalled by 89 North in September 2, 2020. Reason: The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI;…. Check the official notice for the remedy. Verify recall #Z-2820-2020 with the FDA Devices before acting.
The recall
89 North issued this moderate-severity FDA Devices recall-The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI;….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2820-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2820-2020) was formally reported on September 2, 2020, with the manufacturer initiating the action on January 31, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. 89 North is listed as the recalling firm, operating out of Williston, VT. Federal records list the affected scope as 188.
The documented reason for this recall is: The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment o… Distribution data in the federal record shows the product reached: Worldwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
188
Related Recalls
6
3 from same agency
LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science fluorescence imaging
The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2820-2020 |
| Date reported | September 2, 2020 |
| Date initiated | January 31, 2020 |
| Recalling firm | 89 North |
| Firm location | Williston, VT |
| Affected scope | 188 |
| Distribution | Worldwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 2, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.