Severity
Moderate
FDA Devices recall · Reported August 22, 2018
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimm…
Intuitive Surgical, Inc. recalled The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model… - a moderate-severity action.
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model… was recalled by Intuitive Surgical, Inc. in August 22, 2018. Reason: An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may re…. Check the official notice for the remedy. Verify recall #Z-2835-2018 with the FDA Devices before acting.
The recall
Intuitive Surgical, Inc. issued this moderate-severity FDA Devices recall-An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may re….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2835-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2835-2018) was formally reported on August 22, 2018, with the manufacturer initiating the action on May 17, 2018. It is classified under Moderate severity (Class II), with a current status of Ongoing. Intuitive Surgical, Inc. is listed as the recalling firm, operating out of Sunnyvale, CA. Federal records list the affected scope as 1.
The documented reason for this recall is: An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced… Distribution data in the federal record shows the product reached: U.S.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1
Related Recalls
6
3 from same agency
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2835-2018 |
| Date reported | August 22, 2018 |
| Date initiated | May 17, 2018 |
| Recalling firm | Intuitive Surgical, Inc. |
| Firm location | Sunnyvale, CA |
| Affected scope | 1 |
| Distribution | U.S. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 22, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.