Severity
Moderate
FDA Devices recall · Reported September 28, 2016
The handle can separate from the MRx housing due to breakage of mounts on the rear case.
Philips Electronics North America Corporation recalled HeartStart MRx Monitor/Defibrillator Models: M3535A and M3536A - a moderate-severity action.
HeartStart MRx Monitor/Defibrillator Models: M3535A and M3536A was recalled by Philips Electronics North America Corporation in September 28, 2016. Reason: The handle can separate from the MRx housing due to breakage of mounts on the rear case.. Check the official notice for the remedy. Verify recall #Z-2842-2016 with the FDA Devices before acting.
The recall
Philips Electronics North America Corporation issued this moderate-severity FDA Devices recall-The handle can separate from the MRx housing due to breakage of mounts on the rear case..
Sourced from official FDA Devices enforcement records. Verify recall #Z-2842-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2842-2016) was formally reported on September 28, 2016, with the manufacturer initiating the action on August 26, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Electronics North America Corporation is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 394 units.
The documented reason for this recall is: The handle can separate from the MRx housing due to breakage of mounts on the rear case. Distribution data in the federal record shows the product reached: Worldwide distribution. US Nationwide, Canada, AUSTRALIA, FRANCE, GERMANY, INDIA, JAPAN, NETHERLANDS, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, and UNITED KINGDOM.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
394 units
Related Recalls
6
3 from same agency
HeartStart MRx Monitor/Defibrillator Models: M3535A and M3536A
The handle can separate from the MRx housing due to breakage of mounts on the rear case.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2842-2016 |
| Date reported | September 28, 2016 |
| Date initiated | August 26, 2016 |
| Recalling firm | Philips Electronics North America Corporation |
| Firm location | Andover, MA |
| Affected scope | 394 units |
| Distribution | Worldwide distribution. US Nationwide, Canada, AUSTRALIA, FRANCE, GERMANY, INDIA, JAPAN, NETHERLANDS, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, and UNITED KINGDOM. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 28, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.