PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported August 12, 2026

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major sys

Due to a defective circuit board that overheats

Recall #
Z-2857-2026
Affected scope
77
Initiated
June 11, 2026
Verify with FDA Devices →
View my saved recalls

Oxos Medical Inc recalled Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Num… - a moderate-severity action.

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Num… was recalled by Oxos Medical Inc in August 12, 2026. Reason: Due to a defective circuit board that overheats. Check the official notice for the remedy. Verify recall #Z-2857-2026 with the FDA Devices before acting.

The recall

Oxos Medical Inc issued this moderate-severity FDA Devices recall-Due to a defective circuit board that overheats.

Moderate
severity level
77 units
affected scope
Class II
classification
August 12, 2026
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2857-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.

Reason for recall

Due to a defective circuit board that overheats

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-2857-2026
Date reported August 12, 2026
Date initiated June 11, 2026
Recalling firm Oxos Medical Inc
Firm location Atlanta, GA
Affected scope 77
Distribution US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

77 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2857-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.. Recalled by Oxos Medical Inc. Units affected: 77.
Why was this product recalled?
Due to a defective circuit board that overheats
Which agency issued this recall?
This recall was issued by the FDA Devices on August 12, 2026. Severity: Moderate. Recall number: Z-2857-2026.
Where was the recalled product distributed?
Distribution: US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported August 12, 2026. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026