Severity
Moderate
FDA Devices recall · Reported August 29, 2018
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries shou…
Philips Electronics North America Corporation recalled Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 86… - a moderate-severity action.
Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 86… was recalled by Philips Electronics North America Corporation in August 29, 2018. Reason: Philips has received several reports in which monitors with lithium ion batteries that had exceeded their bat…. Check the official notice for the remedy. Verify recall #Z-2895-2018 with the FDA Devices before acting.
The recall
Philips Electronics North America Corporation issued this moderate-severity FDA Devices recall-Philips has received several reports in which monitors with lithium ion batteries that had exceeded their bat….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2895-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2895-2018) was formally reported on August 29, 2018, with the manufacturer initiating the action on July 13, 2018. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips Electronics North America Corporation is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as VM 96,324 Units; VS 110,424 Units; VSV 691 Units.
The documented reason for this recall is: Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon rea… Distribution data in the federal record shows the product reached: Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Camb…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
VM 96,324 Units; VS 110,424 Units; VSV 691 Units
Related Recalls
6
3 from same agency
Philips SureSigns VM 3/4/6/8 Patient Monitors, 863063 863064 863065 863066 863068 863077 863085 863086 863317 863287 863288 863289 The SureSigns VM4, VM6, and VM8 patient monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Standard and optional parameters include: " ECG " Respiration " NBP " S2pO " IBP " CO2 " Temperature
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-2895-2018 |
| Date reported | August 29, 2018 |
| Date initiated | July 13, 2018 |
| Recalling firm | Philips Electronics North America Corporation |
| Firm location | Andover, MA |
| Affected scope | VM 96,324 Units; VS 110,424 Units; VSV 691 Units |
| Distribution | Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Canada, Cambodia, Chile, China, … |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 29, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.