RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.

Reported: September 5, 2018 Initiated: July 16, 2018 #Z-2930-2018

Product Description

RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.

Reason for Recall

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

Details

Units Affected
2357
Distribution
Nationwide distribution. Worldwide foreign distribution.
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
RAPIDLab 1265 Blood Gas Analyzer; 110321852, 10470366, 10491395 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 2357.
Why was this product recalled?
Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 5, 2018. Severity: Moderate. Recall number: Z-2930-2018.