Severity
Moderate
FDA Devices recall · Reported September 9, 2020
This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a…
Siemens Medical Solutions USA, Inc. recalled Symbia S-Series and Symbia T-Series - Product Usage: Intended use of the device SPECT: … - a moderate-severity action.
Symbia S-Series and Symbia T-Series - Product Usage: Intended use of the device SPECT: … was recalled by Siemens Medical Solutions USA, Inc. in September 9, 2020. Reason: This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended …. Check the official notice for the remedy. Verify recall #Z-2934-2020 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc. issued this moderate-severity FDA Devices recall-This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2934-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2934-2020) was formally reported on September 9, 2020, with the manufacturer initiating the action on August 14, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Medical Solutions USA, Inc. is listed as the recalling firm, operating out of Hoffman Estates, IL. Federal records list the affected scope as 412 units.
The documented reason for this recall is: This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a potential risk. Distribution data in the federal record shows the product reached: US Nationwide distribution including in the states of Alabama, Alaska, Arizona, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massac…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
412 units
Related Recalls
6
3 from same agency
Symbia S-Series and Symbia T-Series - Product Usage: Intended use of the device SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.
This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a potential risk.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2934-2020 |
| Date reported | September 9, 2020 |
| Date initiated | August 14, 2020 |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Firm location | Hoffman Estates, IL |
| Affected scope | 412 units |
| Distribution | US Nationwide distribution including in the states of Alabama, Alaska, Arizona, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, M… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 9, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.