BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.

Reported: September 16, 2020 Initiated: August 4, 2020 #Z-2939-2020

Product Description

BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.

Reason for Recall

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.

Details

Recalling Firm
CareFusion 303, Inc.
Units Affected
248,519
Distribution
Worldwide: US* (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY); United Arab Emirates (AE); Australia (AU); Canada (CA); Hungary (HU); Israel (IL); Japan (JP)* Kuwait (KW); New Zealand (NZ); Philippines (PH); Qatar (QA); Saudi Arabia (SA); Singapore (SG); Taiwan (TW); Turkey (TR)*. *There are 159 U.S. government and/or military consignees including two located in Japan (JP) and Turkey (TR).
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.. Recalled by CareFusion 303, Inc.. Units affected: 248,519.
Why was this product recalled?
Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 16, 2020. Severity: Critical. Recall number: Z-2939-2020.