Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible pre…
Roche Diagnostics Corporation recalled Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre… - a moderate-severity action.
Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre… was recalled by Roche Diagnostics Corporation in September 5, 2018. Reason: Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre…. Check the official notice for the remedy. Verify recall #Z-2944-2018 with the FDA Devices before acting.
The recall
Roche Diagnostics Corporation issued this moderate-severity FDA Devices recall-Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2944-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2944-2018) was formally reported on September 5, 2018, with the manufacturer initiating the action on July 16, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Corporation is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 12 units.
The documented reason for this recall is: Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozz… Distribution data in the federal record shows the product reached: US Distribution to states of: CA, NY,TX, WA, and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
12 units
Related Recalls
6
3 from same agency
Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.
Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2944-2018 |
| Date reported | September 5, 2018 |
| Date initiated | July 16, 2018 |
| Recalling firm | Roche Diagnostics Corporation |
| Firm location | Indianapolis, IN |
| Affected scope | 12 units |
| Distribution | US Distribution to states of: CA, NY,TX, WA, and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 5, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.