Severity
Moderate
FDA Devices recall · Reported September 23, 2020
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Addit…
Randox Laboratories Ltd. recalled Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended f… - a moderate-severity action.
Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended f… was recalled by Randox Laboratories Ltd. in September 23, 2020. Reason: An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover a…. Check the official notice for the remedy. Verify recall #Z-2963-2020 with the FDA Devices before acting.
The recall
Randox Laboratories Ltd. issued this moderate-severity FDA Devices recall-An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover a….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2963-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2963-2020) was formally reported on September 23, 2020, with the manufacturer initiating the action on August 11, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Randox Laboratories Ltd. is listed as the recalling firm, operating out of Crumlin (North), N/A. Federal records list the affected scope as 4.
The documented reason for this recall is: An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented … Distribution data in the federal record shows the product reached: US Nationwide distribution including in the states of WV, and Puerto Rico.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4
Related Recalls
6
3 from same agency
Lactate Dehydrognease P-L, (UV), Catalog Number LD3818 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2963-2020 |
| Date reported | September 23, 2020 |
| Date initiated | August 11, 2020 |
| Recalling firm | Randox Laboratories Ltd. |
| Firm location | Crumlin (North), N/A |
| Affected scope | 4 |
| Distribution | US Nationwide distribution including in the states of WV, and Puerto Rico. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 23, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.