Severity
Moderate
FDA Devices recall · Reported September 23, 2020
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed…
Inpeco S.a. recalled FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-… - a moderate-severity action.
FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-… was recalled by Inpeco S.a. in September 23, 2020. Reason: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potenti…. Check the official notice for the remedy. Verify recall #Z-2973-2020 with the FDA Devices before acting.
The recall
Inpeco S.a. issued this moderate-severity FDA Devices recall-Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potenti….
Sourced from official FDA Devices enforcement records. Verify recall #Z-2973-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-2973-2020) was formally reported on September 23, 2020, with the manufacturer initiating the action on August 7, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Inpeco S.a. is listed as the recalling firm, operating out of Lugano, N/A. Federal records list the affected scope as 2 US and 6 OUS.
The documented reason for this recall is: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERL…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
2 US and 6 OUS
Related Recalls
6
0 from same agency
FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory Automation system - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-2973-2020 |
| Date reported | September 23, 2020 |
| Date initiated | August 7, 2020 |
| Recalling firm | Inpeco S.a. |
| Firm location | Lugano, N/A |
| Affected scope | 2 US and 6 OUS |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA, filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported September 23, 2020.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.