FDA Devices recall · Reported September 23, 2020

Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed…

Recall #
Z-2974-2020
Affected scope
2 US and 6 OUS
Initiated
August 7, 2020
Verify with FDA Devices →
View my saved recalls

Inpeco S.a. recalled Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-29… - a moderate-severity action.

Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-29… was recalled by Inpeco S.a. in September 23, 2020. Reason: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potenti…. Check the official notice for the remedy. Verify recall #Z-2974-2020 with the FDA Devices before acting.

The recall

Inpeco S.a. issued this moderate-severity FDA Devices recall-Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potenti….

Moderate
severity level
Class II
classification
September 23, 2020
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-2974-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.

Reason for recall

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-2974-2020
Date reported September 23, 2020
Date initiated August 7, 2020
Recalling firm Inpeco S.a.
Firm location Lugano, N/A
Affected scope 2 US and 6 OUS
Distribution Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-2974-2020) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Accelerator a3600 system with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.. Recalled by Inpeco S.a.. Units affected: 2 US and 6 OUS.
Why was this product recalled?
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Which agency issued this recall?
This recall was issued by the FDA Devices on September 23, 2020. Severity: Moderate. Recall number: Z-2974-2020.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported September 23, 2020. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026