Perceval Dual Collapser, Size S/M, REF: ICV1235, Sterile EO, Rx Only, CE, UDI: 08022057014832 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.

Reported: September 30, 2020 Initiated: July 29, 2020 #Z-2988-2020

Product Description

Perceval Dual Collapser, Size S/M, REF: ICV1235, Sterile EO, Rx Only, CE, UDI: 08022057014832 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.

Reason for Recall

LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.

Details

Recalling Firm
Sorin Group Italia S.r.l.
Units Affected
498 units (component in kits 1-4)
Distribution
US Nationwide distribution including in the states of CA, FL, ID, ME, MN, MO, MS, NC, SD, OH, OR, PA, SD, TN, TX, VA, WA, WI and the countries of Canada, Japan, Germany, Netherland, Switzerland, Sweden, UK, Spain, Italy, France, Belgium, Finland, Greece, Slovenia, Malta, Poland, Romania, Croatia, Hungary, Serbia, Australia, Israel, Georgia, Iran, Kuwait, Oman, Quatar, Singapore, Thailand, Taiwan, Russia, South Africa.
Location
Saluggia, N/A

Frequently Asked Questions

What product was recalled?
Perceval Dual Collapser, Size S/M, REF: ICV1235, Sterile EO, Rx Only, CE, UDI: 08022057014832 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.. Recalled by Sorin Group Italia S.r.l.. Units affected: 498 units (component in kits 1-4).
Why was this product recalled?
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 30, 2020. Severity: Moderate. Recall number: Z-2988-2020.