Product Description

INFORM HPV III Fam 16 Probe - US Export, Catalog Number 05278856001, model 800-4295 Immunohistochemistry (IHC) for in vitro diagnostic use.

Reason for Recall

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Details

Units Affected
195 units
Distribution
worldwide
Location
Oro Valley, AZ

Frequently Asked Questions

What product was recalled?
INFORM HPV III Fam 16 Probe - US Export, Catalog Number 05278856001, model 800-4295 Immunohistochemistry (IHC) for in vitro diagnostic use.. Recalled by Ventana Medical Systems Inc. Units affected: 195 units.
Why was this product recalled?
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.