AnastoClip GC Closure System, REF# 4007-08 a closure system designed to penetrate the outer layers of tissue during the forming of an anastomosis or closing of a vessel with possible intraluminal penetration.
Product Description
AnastoClip GC Closure System, REF# 4007-08 a closure system designed to penetrate the outer layers of tissue during the forming of an anastomosis or closing of a vessel with possible intraluminal penetration.
Reason for Recall
The wrong instructions for use (IFU) were supplied with the AnastoClip GC closure system. The IFU provided lists some indications which are not cleared by the FDA for devices distributed in the United States: 1) closure of arteriotomies and veinotomies, 2) attachment of synthetic vascular prostheses, and 3) approximation of dural tissue.
Details
- Recalling Firm
- LeMaitre Vascular, Inc.
- Units Affected
- N/A
- Distribution
- Distributed to accounts in FL, NV, OK, SC, and TX.
- Location
- Burlington, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11