PlainRecalls
FDA Devices Moderate Class II Terminated

AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLab

Reported: October 7, 2020 Initiated: September 8, 2020 #Z-3052-2020

Product Description

AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005)

Reason for Recall

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Details

Recalling Firm
ESAOTE S.P.A
Units Affected
989 total adapters in the US
Distribution
US Nationwide
Location
Genova, N/A

Frequently Asked Questions

What product was recalled?
AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable ultrasound diagnostic systems: The AC/DC adapter is intended for use with the following portable ultrasound systems: MyLabAlpha (101740000) MyLabGamma (101741000) MyLabOmega (101740004) MyLabSigma (101741004) MyLabSigma elite (101741007) MyLabAlphaVET (101740001) MyLabGammaVET (101741001) MyLabDelta (100743000) MyLabDeltaVET (100743001) MyLabOmegaVET (101740005) MyLabSigmaVET (101741005). Recalled by ESAOTE S.P.A. Units affected: 989 total adapters in the US.
Why was this product recalled?
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 7, 2020. Severity: Moderate. Recall number: Z-3052-2020.