PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported August 30, 2017

Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solutio

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Recall #
Z-3058-2017
Affected scope
6 kits
Initiated
June 29, 2017
Verify with FDA Devices →
View my saved recalls

Medtronic Navigation, Inc. recalled Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908… - a moderate-severity action.

Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908… was recalled by Medtronic Navigation, Inc. in August 30, 2017. Reason: The instruments are not able to pass instrument verification which is performed prior to the use of each inst…. Check the official notice for the remedy. Verify recall #Z-3058-2017 with the FDA Devices before acting.

The recall

Medtronic Navigation, Inc. issued this moderate-severity FDA Devices recall-The instruments are not able to pass instrument verification which is performed prior to the use of each inst….

Moderate
severity level
Class II
classification
August 30, 2017
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-3058-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it

Reason for recall

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-3058-2017
Date reported August 30, 2017
Date initiated June 29, 2017
Recalling firm Medtronic Navigation, Inc.
Firm location Louisville, CO
Affected scope 6 kits
Distribution Worldwide Distribution - US (nationwide) and Internationally to Canada.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-3058-2017) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it. Recalled by Medtronic Navigation, Inc.. Units affected: 6 kits.
Why was this product recalled?
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 30, 2017. Severity: Moderate. Recall number: Z-3058-2017.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US (nationwide) and Internationally to Canada..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported August 30, 2017. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026