Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.

Reported: September 18, 2024 Initiated: July 10, 2024 #Z-3062-2024

Product Description

Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.

Reason for Recall

Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.

Details

Units Affected
625 units
Distribution
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
Location
Tempe, AZ

Frequently Asked Questions

What product was recalled?
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.. Recalled by Bard Peripheral Vascular Inc. Units affected: 625 units.
Why was this product recalled?
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 18, 2024. Severity: Moderate. Recall number: Z-3062-2024.