Severity
Moderate
FDA Devices recall · Reported August 30, 2017
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning…
Pentax of America Inc recalled Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in … - a moderate-severity action.
Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in … was recalled by Pentax of America Inc in August 30, 2017. Reason: Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ul…. Check the official notice for the remedy. Verify recall #Z-3068-2017 with the FDA Devices before acting.
The recall
Pentax of America Inc issued this moderate-severity FDA Devices recall-Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ul….
Sourced from official FDA Devices enforcement records. Verify recall #Z-3068-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-3068-2017) was formally reported on August 30, 2017, with the manufacturer initiating the action on May 23, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Pentax of America Inc is listed as the recalling firm, operating out of Montvale, NJ. Federal records list the affected scope as US: 116 units.
The documented reason for this recall is: Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solut… Distribution data in the federal record shows the product reached: Nationwide. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
US: 116 units
Related Recalls
6
0 from same agency
Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-3068-2017 |
| Date reported | August 30, 2017 |
| Date initiated | May 23, 2017 |
| Recalling firm | Pentax of America Inc |
| Firm location | Montvale, NJ |
| Affected scope | US: 116 units |
| Distribution | Nationwide |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 30, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.