Severity
Moderate
FDA Devices recall · Reported September 6, 2017
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failu…
Siemens Medical Solutions USA, Inc recalled Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.z… - a moderate-severity action.
Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.z… was recalled by Siemens Medical Solutions USA, Inc in September 6, 2017. Reason: Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coola…. Check the official notice for the remedy. Verify recall #Z-3098-2017 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc issued this moderate-severity FDA Devices recall-Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coola….
Sourced from official FDA Devices enforcement records. Verify recall #Z-3098-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-3098-2017) was formally reported on September 6, 2017, with the manufacturer initiating the action on February 18, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Medical Solutions USA, Inc is listed as the recalling firm, operating out of Malvern, PA. Federal records list the affected scope as 986 units in USA.
The documented reason for this recall is: Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in… Distribution data in the federal record shows the product reached: Nationwide Distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
986 units in USA
Related Recalls
6
0 from same agency
Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-3098-2017 |
| Date reported | September 6, 2017 |
| Date initiated | February 18, 2016 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Firm location | Malvern, PA |
| Affected scope | 986 units in USA |
| Distribution | Nationwide Distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 6, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.