Severity
Moderate
FDA Devices recall · Reported September 18, 2024
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample…
Siemens Healthcare Diagnostics, Inc. recalled Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of … - a moderate-severity action.
Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of … was recalled by Siemens Healthcare Diagnostics, Inc. in September 18, 2024. Reason: Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous…. Check the official notice for the remedy. Verify recall #Z-3100-2024 with the FDA Devices before acting.
The recall
Siemens Healthcare Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous….
Sourced from official FDA Devices enforcement records. Verify recall #Z-3100-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-3100-2024) was formally reported on September 18, 2024, with the manufacturer initiating the action on July 26, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Siemens Healthcare Diagnostics, Inc. is listed as the recalling firm, operating out of East Walpole, MA. The number of affected units is not disclosed in the agency filing, which is common for drug and food recalls where lot-level tracking supersedes unit counts.
The documented reason for this recall is: Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not pr… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
Unknown
Related Recalls
6
3 from same agency
Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995490
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-3100-2024 |
| Date reported | September 18, 2024 |
| Date initiated | July 26, 2024 |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Firm location | East Walpole, MA |
| Affected scope | Not disclosed |
| Distribution | Worldwide - US Nationwide distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 18, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.