The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Reported: October 2, 2024 Initiated: November 23, 2023 #Z-3158-2024

Product Description

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Reason for Recall

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

Details

Units Affected
130
Distribution
US Nationwide and Worldwide
Location
Lexington, MA

Frequently Asked Questions

What product was recalled?
The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.. Recalled by FUJIFILM Healthcare Americas Corporation. Units affected: 130.
Why was this product recalled?
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 2, 2024. Severity: Moderate. Recall number: Z-3158-2024.