Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143

Reported: October 2, 2024 Initiated: September 3, 2024 #Z-3193-2024

Product Description

Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143

Reason for Recall

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Details

Recalling Firm
Philips North America Llc
Units Affected
3 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Location
Cambridge, MA

Frequently Asked Questions

What product was recalled?
Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143. Recalled by Philips North America Llc. Units affected: 3 units.
Why was this product recalled?
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 2, 2024. Severity: Moderate. Recall number: Z-3193-2024.