GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.

Reported: October 2, 2024 Initiated: August 5, 2024 #Z-3214-2024

Product Description

GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.

Reason for Recall

GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.

Details

Recalling Firm
GE Healthcare
Units Affected
2138 units
Distribution
Worldwide distribution.
Location
Chicago, IL

Frequently Asked Questions

What product was recalled?
GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.. Recalled by GE Healthcare. Units affected: 2138 units.
Why was this product recalled?
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 2, 2024. Severity: Moderate. Recall number: Z-3214-2024.