FDA Devices recall · Reported October 3, 2018
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following product name sold under: Product Name (Model Number): A. 3.0T MR750 32CH (911000-3); B. 3.0T MR750W (5911000-3, 5481490-2, 5499460-7, 5499460-10, 5373011-3, 5352293, 5911000-3); C. 3.0T MR750W GEM (5499460-10, 5481490-2); D. 3.0T MR750W GEM ENAB (5911000-3); E. 3.0T MR750W GEM ENABLED (5352293, 5481490-2, 5481490-2, 5396469-3, 2394952-100, 5911000-3, 5373011-3); F. 3.0T MR750W GEM
GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get w…
Ge Medical Systems, LLC recalled Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following produ… - a moderate-severity action.
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following produ… was recalled by Ge Medical Systems, LLC in October 3, 2018. Reason: GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic R…. Check the official notice for the remedy. Verify recall #Z-3279-2018 with the FDA Devices before acting.
The recall
Ge Medical Systems, LLC issued this moderate-severity FDA Devices recall-GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic R….
- Moderate
- severity level
- 516 units
- affected scope
- Class II
- classification
- October 3, 2018
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-3279-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following product name sold under: Product Name (Model Number): A. 3.0T MR750 32CH (911000-3); B. 3.0T MR750W (5911000-3, 5481490-2, 5499460-7, 5499460-10, 5373011-3, 5352293, 5911000-3); C. 3.0T MR750W GEM (5499460-10, 5481490-2); D. 3.0T MR750W GEM ENAB (5911000-3); E. 3.0T MR750W GEM ENABLED (5352293, 5481490-2, 5481490-2, 5396469-3, 2394952-100, 5911000-3, 5373011-3); F. 3.0T MR750W GEM ENABLED PLU (5352293); G. 3.0T MR750W GEM ENABLED PLUS (5352293, 5499460-10); H. 3.0T MR750W W GEM (5352293, 5911000-3); I. 3.0T MR750W WITH GEM (5481490-2, 5911000-3); J. 3-0T MR750 32 CHANNEL (5352293); K. 3-0T MR750W FUTURE GEM CAPABLE (5352293, 5911000-3, 5499460-7); L. 3-0T MR750W GEM (5495558, 5352293); M. 3-0T MR750W GEM ENABLED (5911000-3, 5352293, 2394952-100, 5481490-2, 5499460-7, 5373011-3, 5495558, 5481490, 5396469-3, 5481490-3); N. 3-0T MR750W GEM ENABLED PLUS (5911000-3, 5352293, 5499460-10, 5481490-2); O. 3-0T MR750W NON GEM (5911000-3); P. 3T 750W (5921000-2, 5911000-3, 5396469-3, 5352293, 5481490-2, 5481490-3); Q. DISCOVERY MR 750W 3.0T (5481490-2, 5911000-3, 5499460-10); R. DISCOVERY MR750W (5911000-3); S. DISCOVERY MR750W 3 (5481490-2, 5911000-3); T. DISCOVERY MR750W 3T (5911000-3, 5396469-3, 5373011-3, 5481490-2); U. DISCOVERY MR750W GEM (5396469-3, 5352293); V. DISCOVERY MR750W WITH GEM (5911000-3); W. MR 3.0T MR750 W (5352293); X. MR 3.0T MR750W GEM ENABLED (5911000-3, 5352293); Y. MR 3-0T MR750W (5911000-3); Z. MR DISCOVERY 750W GEM (5911000-3); AA. MR OPTIMA MR750W 3.0T (5911000-3); BB. MR750W (5911000-3); CC. MR750W GEM ENABLED (5911000-3, 5499460-7); DD. MR750W GEM ENABLED PLUS (5352293); EE. MR750W PGR 32-CHANNEL (5911000-3); FF. 3.0T MR750W GEM ENAB (5911000-3) Product Usage: The Discovery MR750w 3.0T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750w 3.0T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance..
Reason for recall
GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-3279-2018 |
| Date reported | October 3, 2018 |
| Date initiated | June 6, 2018 |
| Recalling firm | Ge Medical Systems, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 516 Units Total |
| Distribution | Worldwide Distribution - US Nationwide and in the states of AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, US, UT, VA, WA, WI, and WY. and the coun… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-3279-2018) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026