Severity
Moderate
CPSC recall · Reported August 1, 2002
An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccur…
Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. recalled This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are d… - a moderate-severity action.
CPSC, Greenlee Textron Inc. Announce Recall of Electrical Testing Meters was recalled by Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. in August 1, 2002. Reason: An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to…. Remedy: Consumers should stop using the recalled meters immediately and contact Greenlee to recei…. Verify recall #02218 with the CPSC before acting.
The recall
Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. issued this moderate-severity CPSC recall-An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to….
Sourced from official CPSC enforcement records. Verify recall #02218 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This CPSC action (record #02218) was formally reported on August 1, 2002. It is classified under Moderate severity, with a current status of Active. Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. is listed as the recalling firm. Federal records list the affected scope as About 650.
The documented reason for this recall is: An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current … Distribution information was not included in the agency filing, so consumers should assume broad potential exposure until the firm publishes point-of-sale details. The remedy documented by the agency is: Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter. - consumers holding this product should act on that instruction rather than relying on general guidance.
Within the same product category the archive holds 6 closely related recalls, 6 from CPSC - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 609 electronics recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
About 650
Related Recalls
6
6 from same agency
This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are dark green with bright yellow features. The model number and the words "Greenlee Test Instruments" are printed across the front of the meter. A silver plate on the back of the meter displays the serial number (S/N). Serial numbers included in the recall range from 0203540001 through 0203540650. The electrical meters were made in Taiwan. CM-700 and CM-750 meters outside the above serial number range are not included in this recall.
An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current creates the potential for electric shock or an electrocution hazard from the equipment being tested.
Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter.
| Attribute | Value |
|---|---|
| Agency | U.S. Consumer Product Safety Commission |
| Severity class | Moderate |
| Status | Active |
| Recall number | 02218 |
| Date reported | August 1, 2002 |
| Date initiated | August 1, 2002 |
| Recalling firm | Greenlee Textron, Inc. (Greenlee), of Rockford, Ill. |
| Affected scope | About 650 |
| Distribution | Not disclosed |
| Official source | CPSC notice → |
Profile values are sourced directly from the official CPSC enforcement record. Source: U.S. Consumer Product Safety Commission.
What to do with this recall
Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter.
This page summarizes the official CPSC record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Shenzhen Bopu Product Design Co., Ltd., dba POPOOO, of China · 2026-07-02
· 2026-06-25
· 2026-06-25
· 2026-06-25
School Specialty, LLC, of Greenville, Wisconsin · 2026-06-11
Compare this recall with Projecting LED Finger Light Toys Recalled Due to Risk of Se… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: CPSC, reported August 1, 2002. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.