CPSC, Greenlee Textron Inc. Announce Recall of Electrical Testing Meters

Reported: August 1, 2002 Initiated: August 1, 2002 #02218

Product Description

This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are dark green with bright yellow features. The model number and the words "Greenlee Test Instruments" are printed across the front of the meter. A silver plate on the back of the meter displays the serial number (S/N). Serial numbers included in the recall range from 0203540001 through 0203540650. The electrical meters were made in Taiwan. CM-700 and CM-750 meters outside the above serial number range are not included in this recall.

Reason for Recall

An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current creates the potential for electric shock or an electrocution hazard from the equipment being tested.

Remedy

Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter.

Frequently Asked Questions

What product was recalled?
This recall involves certain CM-700 and CM-750 electrical meters. The 8-inch meters are dark green with bright yellow features. The model number and the words "Greenlee Test Instruments" are printed across the front of the meter. A silver plate on the back of the meter displays the serial number (S/N). Serial numbers included in the recall range from 0203540001 through 0203540650. The electrical meters were made in Taiwan. CM-700 and CM-750 meters outside the above serial number range are not included in this recall.. Recalled by Greenlee Textron, Inc. (Greenlee), of Rockford, Ill.. Units affected: About 650.
Why was this product recalled?
An incompatible grommet, located in the battery compartment to protect internal wires, can cause the meter to provide inaccurate voltage and current readings. A meter that inaccurately indicates zero voltage or current creates the potential for electric shock or an electrocution hazard from the equipment being tested.
What should consumers do?
Consumers should stop using the recalled meters immediately and contact Greenlee to receive a replacement meter.
Which agency issued this recall?
This recall was issued by the CPSC on August 1, 2002. Severity: Moderate. Recall number: 02218.