FDA Devices recall · Reported October 11, 2023
Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry System (Refurbished) (Product Code 6801890), VITROS 3600 Immunodiagnostic System (Product Code 6802783), VITROS 3600 Immunodiagnostic System (Refurbished) (Product Code 6802914), VITROS 4600 Chemistry System (Product Code 6802445), VITROS 4600 Chemistry System (Refurbished) (Product Code 6900440), VITROS 5600 Integrated System (Product Code 6802413), VITROS 5600 Integrat
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may…
Ortho-Clinical Diagnostics, Inc. recalled Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Produc… - a moderate-severity action.
Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Produc… was recalled by Ortho-Clinical Diagnostics, Inc. in October 11, 2023. Reason: Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 0…. Check the official notice for the remedy. Verify recall #Z-0008-2024 with the FDA Devices before acting.
The recall
Ortho-Clinical Diagnostics, Inc. issued this moderate-severity FDA Devices recall-Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 0….
- Moderate
- severity level
- Class II
- classification
- October 11, 2023
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0008-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry System (Refurbished) (Product Code 6801890), VITROS 3600 Immunodiagnostic System (Product Code 6802783), VITROS 3600 Immunodiagnostic System (Refurbished) (Product Code 6802914), VITROS 4600 Chemistry System (Product Code 6802445), VITROS 4600 Chemistry System (Refurbished) (Product Code 6900440), VITROS 5600 Integrated System (Product Code 6802413), VITROS 5600 Integrated System (Refurbished) (Product Code 6802915), VITROS XT 3400 Chemistry System (Product Code 6844458), and VITROS XT 7600 Integrated System (Product Code 6844461)
Reason for recall
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause VITROS VersaTips and/or VITROS MicroTips to be improperly seated on the proboscis, which may result in falsely elevated Gentamicin and Valproic Acid assay results as well as an increased frequency of condition codes resulting in delayed patient results. The issue was identified by internal testing when reported instruments failed to pass release testing and were traced to the molded Proboscis part J56639 with date codes from 040722 onward.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0008-2024 |
| Date reported | October 11, 2023 |
| Date initiated | August 9, 2023 |
| Recalling firm | Ortho-Clinical Diagnostics, Inc. |
| Firm location | Rochester, NY |
| Affected scope | 9,860 assemblies |
| Distribution | Worldwide - US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0008-2024) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls