Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN, H3100TT, H3100TR, H3100TW, H3100WZ, H2401MP, H2401MN, H2401MT, H2401MR, H2401MW, H2401NJ. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce cross-sectional images through computer reconstruction of the data. The Brivo NM 615 system is intended fo

GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the c…

Recall #
Z-0189-2015
Affected scope
22
Initiated
October 14, 2014
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Ge Healthcare, LLC recalled Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models… - a moderate-severity action.

Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models… was recalled by Ge Healthcare, LLC in November 19, 2014. Reason: GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the…. Check the official notice for the remedy. Verify recall #Z-0189-2015 with the FDA Devices before acting.

The recall

Ge Healthcare, LLC issued this moderate-severity FDA Devices recall-GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the….

Moderate
severity level
22 units
affected scope
Class II
classification
November 19, 2014
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0189-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0189-2015) was formally reported on November 19, 2014, with the manufacturer initiating the action on October 14, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Ge Healthcare, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 22.

The documented reason for this recall is: GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the … Distribution data in the federal record shows the product reached: Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

22

Related Recalls

6

3 from same agency

Product description

Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN, H3100TT, H3100TR, H3100TW, H3100WZ, H2401MP, H2401MN, H2401MT, H2401MR, H2401MW, H2401NJ. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce cross-sectional images through computer reconstruction of the data. The Brivo NM 615 system is intended for General Nuclear Medicine imaging procedures using variety of scanning modes supported by various acquisition types. This generic type of device may include signal analysis and display equipment, patient and equipment supports, and accessories

Reason for recall

GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0189-2015
Date reported November 19, 2014
Date initiated October 14, 2014
Recalling firm Ge Healthcare, LLC
Firm location Waukesha, WI
Affected scope 22
Distribution Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DEN…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

22 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0189-2015) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Brivo NM615, is an all-purpose, single detector integrated nuclear imaging system. Models H3100TP, H3100TN, H3100TT, H3100TR, H3100TW, H3100WZ, H2401MP, H2401MN, H2401MT, H2401MR, H2401MW, H2401NJ. Product Usage: The Brivo NM 615 is an emission computed tomography system intended to detect the location and distribution of gamma ray photon emitting radionuclides in the body and to produce cross-sectional images through computer reconstruction of the data. The Brivo NM 615 system is intended for General Nuclear Medicine imaging procedures using variety of scanning modes supported by various acquisition types. This generic type of device may include signal analysis and display equipment, patient and equipment supports, and accessories. Recalled by Ge Healthcare, LLC. Units affected: 22.
Why was this product recalled?
GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 19, 2014. Severity: Moderate. Recall number: Z-0189-2015.
Where was the recalled product distributed?
Distribution: Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, GREECE, GUADELOUPE, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA,MARTINIQUE, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN,SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, URUGUAY, VENEZUELA, VIET NAAM..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0189-2015) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 19, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.