Severity
Moderate
FDA Devices recall · Reported October 31, 2018
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occu…
Ge Medical Systems, LLC recalled GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2. - a moderate-severity action.
GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2. was recalled by Ge Medical Systems, LLC in October 31, 2018. Reason: GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used wi…. Check the official notice for the remedy. Verify recall #Z-0201-2019 with the FDA Devices before acting.
The recall
Ge Medical Systems, LLC issued this moderate-severity FDA Devices recall-GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used wi….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0201-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0201-2019) was formally reported on October 31, 2018, with the manufacturer initiating the action on December 7, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Ge Medical Systems, LLC is listed as the recalling firm, operating out of Waukesha, WI. Federal records list the affected scope as 356 (98 US; 258 OUS).
The documented reason for this recall is: GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setu… Distribution data in the federal record shows the product reached: Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBL…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,422 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
356 (98 US; 258 OUS)
Related Recalls
6
0 from same agency
GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0201-2019 |
| Date reported | October 31, 2018 |
| Date initiated | December 7, 2015 |
| Recalling firm | Ge Medical Systems, LLC |
| Firm location | Waukesha, WI |
| Affected scope | 356 (98 US; 258 OUS) |
| Distribution | Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NE… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-07-02
· 2026-07-02
· 2026-06-25
· 2026-06-25
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 31, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.