Fukuda Denshi patient monitor model DS-8100M and DS-8100N Product Usage: Use of the Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is indicated in those situations where observation of one or more of the following parameters on an individual patient may be required. ECG (waveform, heart rate, ST-Level and ventricular arrhythmias), respiration, non-invasive blood pressure (NIBP), pulse rate, arterial oxygen saturation (SpO2), carboxyhemoglobin saturation (SpCO)*, methemoglobin s

Reported: December 13, 2017 Initiated: October 16, 2017 #Z-0203-2018 6227 units units

Fukuda Denshi Co., Ltd. issued this FDA Devices recall on December 13, 2017. Classified as Moderate severity (Class II). Approximately 6227 units units are affected. The recall was issued because: For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Sh…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0203-2018) was formally reported on December 13, 2017, with the manufacturer initiating the action on October 16, 2017. It is classified under Moderate severity (Class II), with a current status of Completed. Fukuda Denshi Co., Ltd. is listed as the recalling firm, operating out of Tokyo, N/A. Federal records indicate 6227 units units are affected.

The documented reason for this recall is: For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion… Distribution data in the federal record shows the product reached: Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 9 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

6227 units

Related Recalls

6

6 from same agency

Product Description

Fukuda Denshi patient monitor model DS-8100M and DS-8100N Product Usage: Use of the Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is indicated in those situations where observation of one or more of the following parameters on an individual patient may be required. ECG (waveform, heart rate, ST-Level and ventricular arrhythmias), respiration, non-invasive blood pressure (NIBP), pulse rate, arterial oxygen saturation (SpO2), carboxyhemoglobin saturation (SpCO)*, methemoglobin saturation (SpMet)*, total hemoglobin concentration (SpHb)*, plethysmograph, temperature, invasive blood pressure (IBP), cardiac output, and carbon dioxide concentration (CO2). *: DS-8100M only The target populations of the system are adult, pediatric and neonatal patients with the exception of the ST segment, arrhythmia analysis, and SpHb, for which the target populations are adult and pediatric excluding neonates. These observations can include an audible and visual alarm if any of these parameters exceed values that are established by the clinician. The observations may include the individual or comparative trending of one or more of these parameters over a period of up to 24 hours. The DS-8100N/8100M Patient Monitor is indicated in situations where an instantaneous display of waveform, numeric and trended values is desired. The DS-8100N/8100M Patient Monitor is also indicated where a hard copy record of the physiological parameters, the alarms conditions or the trended values may be required.

Reason for Recall

For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term battery backup).

Details

Recalling Firm
Fukuda Denshi Co., Ltd.
Units Affected
6227 units
Distribution
Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries
Location
Tokyo, N/A

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Completed
Recall number Z-0203-2018
Date reported December 13, 2017
Date initiated October 16, 2017
Recalling firm Fukuda Denshi Co., Ltd.
Units affected 6227 units
Distribution Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

6227 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Fukuda Denshi patient monitor model DS-8100M and DS-8100N Product Usage: Use of the Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is indicated in those situations where observation of one or more of the following parameters on an individual patient may be required. ECG (waveform, heart rate, ST-Level and ventricular arrhythmias), respiration, non-invasive blood pressure (NIBP), pulse rate, arterial oxygen saturation (SpO2), carboxyhemoglobin saturation (SpCO)*, methemoglobin saturation (SpMet)*, total hemoglobin concentration (SpHb)*, plethysmograph, temperature, invasive blood pressure (IBP), cardiac output, and carbon dioxide concentration (CO2). *: DS-8100M only The target populations of the system are adult, pediatric and neonatal patients with the exception of the ST segment, arrhythmia analysis, and SpHb, for which the target populations are adult and pediatric excluding neonates. These observations can include an audible and visual alarm if any of these parameters exceed values that are established by the clinician. The observations may include the individual or comparative trending of one or more of these parameters over a period of up to 24 hours. The DS-8100N/8100M Patient Monitor is indicated in situations where an instantaneous display of waveform, numeric and trended values is desired. The DS-8100N/8100M Patient Monitor is also indicated where a hard copy record of the physiological parameters, the alarms conditions or the trended values may be required.. Recalled by Fukuda Denshi Co., Ltd.. Units affected: 6227 units.
Why was this product recalled?
For the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term battery backup).
Which agency issued this recall?
This recall was issued by the FDA Devices on December 13, 2017. Severity: Moderate. Recall number: Z-0203-2018.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US Nationwide and in the countries of UK, and Japan and via distributors in the other countries.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0203-2018) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).